I. E+H electromagnetic flowmeter panel operation
Note: When you need to enter a password, just enter 4685.
1. Press "E" to enter "GROUP SELECT" (main menu), press "+" to find "system units", press "E" to enter, at this time, the screen displays "unit volume flow", press "+" and "-" to set, press "E" to confirm the storage. After setting, press "+" and "-" at the same time to return to the main menu
2. From the main menu, press "+" to find "user interface", press "E" to enter, it will display "assigne line 1", press "+" to find "volume flow", press "E" to confirm, and return to the main menu.
3. From the main menu, press "+" to find "current output 1" and press "E" to enter. The screen displays "assign current". Press "+" to find "volume flow". Press "E" to confirm. Press "E" again. The screen displays "current span". Press "+" to select "4-20mA hart nam". Press "E" to confirm. The display screen displays "value 0_4mA". Enter the corresponding value of 0-4mA. Press "E" to confirm and enter the next submenu. At this time, the display screen displays "value20mA". Enter the corresponding value of 20mA. Press "E" to confirm and save, and return to the main menu.
4. From the main menu, press "+" to find "process parameter" and press "E" to enter. The screen is "assigne lf-cut off". Press "+" to find "volume flow" and press "E" to confirm. Press "E" again to enter "on-val. Lf-cutoff". Enter the value (this value is 3~5% of the total flow) and press "E" to save.
II. Basic commissioning instructions of E+H electromagnetic flowmeter Promag 50
【Function key description】
1. There are three buttons on the operation interface: "+", "-" and "E";
2. "+" and "-" can be used to switch menus up and down, or to set values in the numeric input interface (when the numeric input interface is shaded, press "+" or "-", the shadow disappears and a single cursor appears, and the current value can be modified, such as the number "1" is changed to "2" by pressing "+" once); "E" is the confirmation key, used to enter the corresponding menu or confirm the value (if you need to enter multiple digits, such as "178", first enter 1 in the first digit, press E to confirm, enter the second digit, enter "7", then press E to enter the third digit, enter "8", and press E to confirm).
3. Press the "+" and "-" keys simultaneously to return to the correct menu. If you accidentally enter the wrong menu during the setting process, press them a few more times to return to the main measurement interface, and then re-enter the correct menu to set.
【Quick Settings】
1. After the meter is powered on, it goes through the startup procedure to enter the measurement mode, and press "E" to enter the function menu:

2. Press "+" or "-" to "QUICK SETUP", and press "E" to enter the quick setup function:

3. Press the "+" or "-" key to enter the password. Press "+" or "-" to enter the password "50" and press "E" to confirm.
And press "+" or "-" again to select "", press "E" to confirm;

4. Press "E" several times to "UNIT VOLUME FLOW", press "+" or "-" to change the instantaneous flow unit.
After completion, press "E" to confirm:

5. Press "E" several times until the display shows "VALUE 20mA", and press "+" or "-" to enter the value corresponding to 20mA:

6. Press "E" several times to return to the function menu to complete the quick setting;
【Process parameter setting】
(small signal removal and empty pipe detection function)
1. Press "E" to enter the function menu:

2. Press the "+" or "-" key to "PROCESSPARAMETER", as shown below:

3. Press "E" several times until the display shows the figure, and press "+" or "-" to modify the small signal cutoff value (the recommended small signal cutoff value is 1% to 5% of the range):

4. Press "E" several times until the display is as shown, press "+" or "-" to select "ON STANDARD", and press "E" to confirm;

5. Press "E" several times to complete the process parameter setting until the display is:

6. Press "+" and "-" several times at the same time to exit the measurement mode.
After the above debugging, the basic settings of the Promag 50 series electromagnetic flowmeter have been completed.